User Maintenance Instructions
1. Introduction
Medical staff or users should follow this schedule to maintain the micomme home-care non-invasive ventilator to ensure its safety and reliability.
2. Maintenance schedule
Implement preventive maintenance and maintenance of the non-invasive ventilator according to the time in the table.
Maintenance schedule
Item
Frequency
Methode
Operator
Ventilator tubes, masks
Daily or on demand
1. Empty the water stains inside the tubes of the ventilator. Keep tubes and masks dry
Medical staff or user
Water chamber
Daily or on demand
1. Empty the water in the water chamber and clean it every day
Filter cotton
Weekly or on demand
1. Clean the air filter cotton with a neutral detergent in warm water
Cover
According to the degree of taint
Wipe the exterior of the equipment with a rag and neutral detergent
2. Check whether the accessories are damaged, and replace them in time if damaged
3. Soak the mask/tube in a commercially available disinfectant, such as 2% glutaraldehyde for 20 minutes. Do not disinfect at high temperature. Rinse the mask/tube with plenty of clean drinking water, and then air-dry it naturally and avoid direct sunlight.
4. It must be disinfected after changing users
5. Replace every 6 to 12 months according to the degree of dirt
2. Disinfect with the mask and tube after changing users
3. Check whether the water chamber is in good condition after each cleaning
2. If the air filter is damaged, replace it